What is the torque value when mounting the lap joint SSiC sensor QP018?

What is the torque value when mounting the lap joint SSiC sensor QP018?

The SSiC lap joint flange is stronger than any material. When fixing it to the stainless steel flange, do not over-tighten the value of the bolts.
The maximum bolt value for M6 is 11,7 Nm.
We recommend to keep the torque value below this value, say 10 Nm or 7,3 lbf.ft
Note these instructions:
- there must be a new packing in between the SiC and the steel flange;
- the sensor must be concentrically mounted in the flange;
- there must be a steel ring between the bolt and the SiC;
- the steel flange must be flat;
- the bolts must be tightened crosswise;
- the bolts may not touch the SiC.

If you need more information, please let us know.

The Arenal Support team.



    • Related Articles

    • What is the maximum pressure of our sensor?

      The maximum pressure of our SSiC sensor (QP015-UDP-SIC-LT) is 80 bar. This is not officially tested, but tested during an indicative test. We could test it on higher pressures on request. For higher pressures, like 600 or 6000 bar, we can create a ...
    • What is the best place to mount the density sensor and the thermal massflow meter?

      Hello, The point of installation is very important for a succesful measurement. In the datasheet you find all information about it. The minimum velocity for the density sensor of the flow shall be at least 1 feet or 30 cm. If you expect a lower ...
    • What is the hardness of the sintered silicon carbide probe

      Our sensor QP015-UDP-SIC is made from sintered silicon carbide (SSiC). Arenal is the only company making ultrasonic sensors with this material. Is is designed to fit in mineral processing plants and dredging and rilling applications. It does not wear ...
    • How to get support from Arenal Support Team?

      Hello, Welcome to the Arenal Service & Support Team in The Netherlands. We are trying to support you the best we can. We are working on a full automatic response system for small questions. When this is implemented (possibly end of February) we ...
    • What to do when the USB stick was lost or fails?

      We have two active solutions, one with a In a Proface HMI, the other with a Cimon HMI. In the ProFace, when the USB stick was lost and could not be found, then insert a new USB stick in the analyzer and press SAVE. Two init files will be created on ...