What to do when the USB stick was lost or fails?

What to do when the USB stick was lost or fails?

The USB sticks that we supply are industrial and they never failed. If you use your own USB stick, note that we did tests on them and they all failed the test. So make sure to create a copy of the content of ours before you use it.
We have three active solutions, one with a In a Proface HMI, the other with a Cimon C7E HMI and the last in the Cimon C7X HMI.
In the ProFace, when the USB stick was lost and could not be found, then insert a new USB stick in the analyzer and press SAVE. Two init files will be created on the USB stick. The file from the probe "114..." can be requested by Arenal.
This file is send to you by email or WhatsApp or Skype. Copy this file to your USB stick and then remove its extension ".dat".
When the analyzer does not contain the settings as abov(for example after a software update), then follow the instructions to set the SG, density, TSS and temperature in the attachment (KWR-7QA03-935).

In the Cimon C7E or C7X you are also able to copy the data from the USB stick to the internal directory of the analyzer, as it is a Windows based HMI resp Linux based.

Anyway, find the 3 files: INITPRM1.csv and INITQA03.csv file and the 114... calibration file.

Copy the files from the USB stick on to its root.

Restart the HMI, go to the Configuration Menu in the analyzer. Press Ultrasonic Density Transmitter. Press Calibration. Select the 114... file. save. 

Check the settings in the enclosure. I hope this helped you finding and solving the issue!

Thanks! The Arenal Support Team
Marius de Vries




    • Related Articles

    • How to connect the QA03 analyzer to your PC

      Connect your analyzer system to your PC by connecting the PC with a normal UTP cable to the HMI. Or, on the base board undo the switch PoW and connect the UTP cable in the RJ45 connector. Or, connect to the VPN router (cost apply). Once connected use ...
    • How to get support from Arenal Support Team?

      Hello, Welcome to the Arenal Service & Support Team in The Netherlands. We are trying to support you the best we can. We are working on a full automatic response system for small questions. When this is implemented (possibly end of February) we ...
    • How to perform a slurry (or field) calibration on the C7X

      This instruction is copied from the User Manual: 5.10 Slurry calibration The conditions in our lab will be different from the conditions in the field. Therefore, an initial calibration is needed. After performing the water calibration: · Apply high ...