We have two active solutions, one with a In a Proface HMI, the other with a Cimon HMI.
In the ProFace, when the USB stick was lost and could not be found, then insert a new USB stick in the analyzer and press SAVE. Two init files will be created on the USB stick. The file from the probe "114..." can be requested by Arenal.
This file is send to you by email or WhatsApp or Skype. Copy this file to your USB stick and then remove its extension ".dat".
When the analyzer does not contain the settings as
above (for example after
a software update), then follow the instructions to set the SG, density, TSS and
temperature in the attachment (KWR-7QA03-935).
In the Cimon, you are able to copy the data from the USB stick to the internal directory of the analyzer, as it is a Windows based HMI.
Plug the USB stick in your laptop.
Connect the laptop by ethernet (LAN) cable in the HMI.
Press the Ethernet icon and check the IP address of the unit. In the case of the "picture 1 (IP adress)" see enclosure, it is 192.168.9.107
Make sure your laptop LAN port is isn the same network as this analyzer.
On your laptop start Windows Explorer.
Type in FTP://192.168.9.107
You are now at the directory of the HMI.
Go to XPanel directory
Copy the files from the USB stick on this locations.
Now finally, go to the Configuartion Menu in the analyzer. Press Ultrasonic Density Transmitter. Press Calibration.
Press Change Sensor Calibration. It will find the 114....csv file you just copied. Select it.
I hope this helped you finding and solving the issue!
Thanks! The Arenal Support Team
Marius de Vries